Aortic perforation by active-fixation atrial pacing lead
Authors:Kashani Amir  Mehdirad Ali  Fredman Carey  Barold S Serge
Institution:St John's Mercy Medical Center, St Louis, Missouri, USA. AHK3@aol.com
Abstract:This report describes perforation of the aorta related to the implantation of an active-fixation atrial pacing lead, a previously undocumented complication of pacemaker implantation. The complication was related to excessive tissue penetration by the screw at the tip of the lead or perforation of the lead body by the positioning stylet during manipulation whereupon the stylet traversed the atrial wall and aortic wall. Perforation of the aorta should be part of the differential diagnosis of cardiac tamponade after pacemaker implantation.
Keywords:pacemaker                        atrial pacing                        cardiac tamponade                        aortic perforation