Sensing Failure in a Tiered Therapy Implantable Cardioverter Defibrillator: Role of Auto Adjustable Gain |
| |
Authors: | ABDUL WASE REA NATALE ANWER A. DHALA SANJAY DESHPANDE JASBIR S. SRA ZALMEN BLANCK CHERYL MAGLIO MOHAMMAD R. JAZAYERI MASOOD AKHTAR |
| |
Affiliation: | Wisconsin Electrophysiology Group, University of Wisconsin-Milwaukee Clinical Campus, Milwaukee Heart Institute of Sinai Samaritan Medal Center, and St. Luke's Medical Center, Milwaukee, Wisconsin |
| |
Abstract: | WASE A., et.al .: Sensing Failure in a Tiered Therapy Implantable Cardioverter Defibrillator: Role of Auto Adjustable Gain . Implantable cardioverter defibrillators have an established role in the management of life-threatening tachyarrhythmias. These devices use sophisticated sensing circuitry to detect and promptly treat a vast majority of these arrhythmias. However, they are not foolproof. We report one case where the device failed to sense every other QRS complex during induced ventricular fibrillation due to marked electrical alterans. Thus, undersensing can be a potentially fatal problem despite the use of auto adjustable gain. |
| |
Keywords: | implantable cardioverter defibrillator auto adjustable gain undersensing |
|