Sensing Failure in a Tiered Therapy Implantable Cardioverter Defibrillator: Role of Auto Adjustable Gain
Authors:ABDUL WASE  REA NATALE  ANWER A. DHALA  SANJAY DESHPANDE  JASBIR S. SRA  ZALMEN BLANCK  CHERYL MAGLIO  MOHAMMAD R. JAZAYERI  MASOOD AKHTAR
Affiliation:Wisconsin Electrophysiology Group, University of Wisconsin-Milwaukee Clinical Campus, Milwaukee Heart Institute of Sinai Samaritan Medal Center, and St. Luke's Medical Center, Milwaukee, Wisconsin
Abstract:WASE A., et.al .: Sensing Failure in a Tiered Therapy Implantable Cardioverter Defibrillator: Role of Auto Adjustable Gain . Implantable cardioverter defibrillators have an established role in the management of life-threatening tachyarrhythmias. These devices use sophisticated sensing circuitry to detect and promptly treat a vast majority of these arrhythmias. However, they are not foolproof. We report one case where the device failed to sense every other QRS complex during induced ventricular fibrillation due to marked electrical alterans. Thus, undersensing can be a potentially fatal problem despite the use of auto adjustable gain.
Keywords:implantable cardioverter defibrillator    auto adjustable gain    undersensing