Simple access to the coronary venous system for left ventricular pacing
Authors:Sayad Dany E  Sawar Asad  Curkovic Vladimir  Gallardo Ignacio  Barold S Serge
Institution:Florida Cardiovascular Institute, Tampa, Florida, USA. dsayad@fciheart.com
Abstract:Implantation of the LV lead for biventricular pacing can be challenging, time consuming, and often requires extensive fluoroscopy time. A conventional diagnostic 5 Fr left Amplatz catheter was used to cannulate the coronary sinus in 15 consecutive patients undergoing implantation of a biventricular pacemaker. When the coronary sinus was cannulated, the proximal end of the Amplatz catheter was cut and the coronary sinus sheath was passed over the Amplatz catheter that was then removed. Coronary sinus cannulation was achieved in all 15 patients with a mean fluoroscopy time of 3.34 +/- 1.9 minutes. Subsequent implantation of a biventricular pacemaker was successful and free of complications in all the 15 patients.
Keywords:biventricular pacing                        left ventricular pacing                        coronary sinus                        left Amplatz catheter                        congestive heart failure