A novel intermittent negative air pressure device ameliorates obstructive sleep apnea syndrome in adults
Authors:Hung  Tzu-Chun  Liu  Tien-Jen  Hsieh  Wen-Yeh  Chen  Bo-Nien  Su  Wen-Ko  Sun  Kuang-Hui  Guilleminault  Christian
Institution:1.Department of Biotechnology and Laboratory Science in Medicine, National Yang-Ming University, Taipei, Taiwan
;2.Department of Otolaryngology Head & Neck Surgery, MacKay Memorial Hospital, Taipei Branch, Taipei, Taiwan
;3.School of Biomedical Engineering, Taipei Medical University, Taipei, Taiwan
;4.Division of Chest Medicine, MacKay Memorial Hospital, Hsinchu Branch, Hsinchu, Taiwan
;5.Department of Radiology, MacKay Memorial Hospital, Hsinchu Branch, Hsinchu, Taiwan
;6.Department of Education and Research, Taipei City Hospital, Taipei, Taiwan
;7.Division of Sleep Medicine, Stanford University, Redwood City, CA, USA
;
Abstract:Purpose

Patients with obstructive sleep apnea syndrome (OSAS) have difficulties in compliance with continuous positive airway pressure (CPAP) and the treatment outcome is heterogeneous. We proposed a proof-of-concept study of a novel intermittent negative air pressure (iNAP®) device for physicians to apply on patients who have failed or refused to use CPAP.

Methods

The iNAP® device retains the tongue and the soft palate in a forward position to decrease airway obstruction. A full nightly usage with the device was evaluated with polysomnography. Subgrouping by baseline apnea–hypopnea index (AHI) and body mass index (BMI) with different treatment response criteria was applied to characterize the responder group of this novel device.

Results

Thirty-five patients were enrolled: age 41.9?±?12.2 years (mean?±?standard deviation), BMI 26.6?±?4.3 kg/m2, AHI 41.4?±?24.3 events/h, and oxygen desaturation index (ODI) 40.9?±?24.4 events/h at baseline. AHI and ODI were significantly decreased (p?<?0.001) by the device. Patients with moderate OSAS, with baseline AHI between 15 to 30 events/h, achieved 64% response rate; and non-obese patients, with BMI below 25 kg/m2, achieved 57% response rate, with response rate defined as 50% reduction in AHI from baseline and treated AHI lower than 20. There were minimal side effects reported.

Conclusions

In a proof-of-concept study, the device attained response to treatment as defined, in more than half of the moderate and non-obese OSAS patients, with minimal side effects.