应用Westgard方法评价决定图判断Sysmex XE5000全血细胞分析仪性能 |
| |
引用本文: | 曾建明,龙一飞,李沫,陈中华,邓光远,陈茶.应用Westgard方法评价决定图判断Sysmex XE5000全血细胞分析仪性能[J].实验与检验医学,2012,30(3):225-226,264. |
| |
作者姓名: | 曾建明 龙一飞 李沫 陈中华 邓光远 陈茶 |
| |
作者单位: | 广东省中医院检验科,广东广州,510120 |
| |
基金项目: | 全国重点地区环境与健康专项调查“广东试点”资金资助(编号:21111011101) |
| |
摘 要: | 目的应用Westgard方法评价决定图判断Sysmex XE5000型全血细胞分析仪系统性能的可接受性。方法以实验室室内质控结果的变异系数(CV%)反应方法的不精密度,以实验室参加卫生部临床检验中心2012年室间质评的偏倚(Bias%)反映方法的不准确度,在Westgard方法评价决定图上绘制操作点,并判断各项目的方法性能的可接受性。结果WBC、RBC、HGB、HCT、MCH、MCV、MCHC和PLT的CV%值分别为:1.89、0.6、0.6、0.79、0.9、0.92、1.3和3.2,偏倚分别为-5.71、-2.55、-1.46、-0.27、2.23、0.46、0.27和-6.97。结论采用Westgard方法评价决定图能方便地评价检测仪器的检测性能,Sysmex XE5000全血细胞分析仪主要检测指标性能优秀。
|
关 键 词: | 全血细胞分析仪 质量控制 Westgard方法评价决定图 |
Application of Westgard method evaluation decision chart(MEDx) for judging method performance of Sysmex XE5000 hematologic analyzer |
| |
Institution: | ZENG Jianming,LONE Yifei,LI Mo,et al.Department of Laboratory Medicine,Guangdong Provincial Hospital of Traditional Chinese Medicine,Guangzhou 510120,China |
| |
Abstract: | Objective To evaluate the performance of Sysmex XE5000 hematology analyzer by Westgard method evaluation decision chart(MEDx).Methods The percentage coefficient of variation(CV%) accquired from laboratory internal quality control and bias% from external quality assurance(EQA) were taken as measurement imprecision and inaccuracy respectively for analyzer.Westgard method evaluation decision chart(MEDx) is a graphical tool that provides an easy way to judge a method’s performance.MEDx chart took CV % and bias% as operating point’s x and y-coordinate.Operating points of parameters WBC,RBC,HGB,MCV,MCH,MCHC and PLT in MEDx chart were ploted to judge the performance of hematology analyzer XE5000.Results The CV% of WBC,RBC,HGB,HCT,MCH,MCV,MCHC and PLT were 1.89,0.6,0.6,0.79,0.9,0.92,1.3 and 3.2 respectively,and the bias% of those parameters were-5.71,-2.55,-1.46,-0.27,2.23,0.46,0.27 and-6.97 respectively.All the operating points droped in excellent performance region in MEDx.Conclusions Objective decisions on the performance of methods can be made quickly and easily with the aid of a MEDx,Sysmex XE5000 hematology analyzer possesses excellent performance on WBC,RBC,HGB,HCT,MCV,MCH,MCHC and PLT. |
| |
Keywords: | Hematology analyzer Quality control Westgard method evaluation decision chart(MEDx) |