Factors Affecting Impedance of Percutaneous Leads in Spinal Cord Stimulation
Authors:Kenneth Al  Clayton Varga  Elliot Krames  Joshua Prager  Jan Holsheimer  Ljubomir Manola  Kerry Bradley
Institution:Kenneth Alò,Clayton Varga,Elliot Krames,Joshua Prager,Jan Holsheimer,Ljubomir Manola,Kerry Bradley
Abstract:Objectives. Although the load impedance of a pulse generator has a significant effect on battery life, the electrical impedance of contact arrays in spinal cord stimulation (SCS) has not been extensively studied. We sought to characterize the typical impedance values measured from common quadripolar percutaneous SCS contact arrays. Methods. In 36 patients undergoing percutaneous trial stimulation for various chronic pain conditions, bipolar impedance between adjacent contacts of 64 leads with 9 mm center‐to‐center spacing was measured in two different vertebral level regions, cervical (C3–C7) and lower‐thoracic (T7–T12). Multiple linear regression was applied to analyze the contribution of six variables to the biological tissue portion of the impedance (excluding the resistance of the lead wires). Results. The median impedance in the cervical region (351 ± 90 Ω) was significantly lower (36%, p < 0.001) than in the lower‐thoracic region (547 ± 151 Ω). In addition, time since implant had a weaker but still significant effect on tissue impedance. Conclusions. Results from finite‐difference mathematical modeling of SCS suggest that the difference in tissue impedance related to vertebral level may be due to the dorsoventral position of the lead in the epidural space. The presence of a larger space between the triangularly shaped dorsal part of the vertebral arch and the round shape of the dural sac in the lower‐thoracic region increases the likelihood that the stimulating lead will not make dural contact, and thus “see” an increased impedance from the surrounding epidural fat. This implies that the energy requirements for stimulation in the thoracic region will be higher than in the cervical region, at least during the acute phase of implant.
Keywords:electrical impedance  energy consumption  epidural space  percutaneous lead  spinal cord stimulation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号