首页 | 本学科首页   官方微博 | 高级检索  
检索        


Automatic read-out device for ultraviolet-radiation polymer-film dosimeters
Authors:M G Pepper  B L Diffey
Institution:(1) Electronics Laboratories, University of Kent, Canterbury, Kent, England;(2) Medical Physics Department, Nuffield Ward, Dryburn Hospital Durham DH1 5TW, England
Abstract:An automatic device for measuring the optical absorbance of polymer film used as an ultravioletradiation dosimeter is described. The automated mechanics are based on a commercial slide projector. The ultraviolet radiation is generated by a tungsten quartz-halogen bulb. The optical absorbance is determined by analogue logarithmic circuitry and the resultant value displayed on a digital panel meter. A digital printer or paper-tape punch may be used for hard-copy output. An uncertainty in measurement of 1% is obtained. The device temperature stability is such that this uncertainty can be maintained for analogue circuit temperatures of 20–40°C.
Keywords:Automatic read-out device  Logarithmic  Polymer film  Ultraviolet radiation dosimetry
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏