Influence of Internal Current and Pacing Current on Pacemaker Longevity |
| |
Authors: | ANDREAS SCHUCHERT KARL-HEINZ KUCK |
| |
Affiliation: | Medical Clinic, Department of Cardiology, University-Hospital Eppendurf, Hamburg, Germany |
| |
Abstract: | The effects of lower pulse amplitude on battery current and pacemaker longevity were studied comparing the new, small-sized VVI pacemaker, Minix 8341, with the former model, Pasys 8329. Battery current was telemetrically measured at 0.8. 1.6, 2.5, and 5.0 V pulse amplitude and 0.05, 0.25, 0.5, and 1.0 msec pulse duration. Internal current was assumed to be equal to the battery current at 0.8 V and 0.05 msec. Pacing current was calculated subtracting internal current from battery current. The Minix pacemaker had a significantly lower battery current because of a lower internal current (Minix: 4.1 ±0.1 μA; Pasys: 16.1 ± 0.1 μA); pacing current of both units was similar. At 0.5 msec pulse duration, the programming from 5.0-2.5 V puise amplitude resulted in a greater relative reduction of battery current in the newer pacemaker (51% vs 25%). Projected longevity of each pacemaker was 7.9 years at 5.0 V and 0.5 msec. The programming from 5.0–2.5 V extended the projected longevity by 2.3 years (Pasys) and by 7.1 years (Minix). The longevity was negligibly longer after programming to 1.6 V. Conclusion: Extension of pacemaker longevity can be achieved with the programming to 2.5 V or less if the connected pacemakers need a low internal current for their circuitry. |
| |
Keywords: | pacemaker programming battery current internal and pacing current pacemaker battery capacity |
|
|