Measurement of area of surplus X-ray radiation by dose-area product meter at image intensifier in angiography X-ray units
Authors:Seguchi Shigenobu  Saijou Takaya  Nagao Tatsuji  Ishikawa Yoshinobu  Mizuno Shouta  Nakamura Akirou
Affiliation:Department of Radiology, Nagoya Daini Red Cross Hospital.
Abstract:The aim of this study was to assess the area of surplus X-ray radiation (surplus area) of three image intensifiers (I.I.) of two angiography X-ray units. Surplus area refers to an area surrounded with a circular collimator and effective field of view (FOV). Dose-area product (DAP) was used to measure surplus X-rays. A ratio of surplus area when based on each FOV showed an 8% to 39.3% increase with expansion of FOV, surplus area showed an 18 cm(2) to 95 cm(2) decrease with FOV. When the circular collimator was narrowed to a degree that could be confirmed partially in effective FOV, the rate of surplus area was decreased almost as much as the rate of scattered radiation. Surface dose was almost unchanged. The radiation exposure of the physician as well as that of the patient was reduced by the reduction of surplus area. The DAP meter, while a tool for patient radiation exposure management, is also highly useful for machinery management such as the measurement of surplus area at the I.I. system.