Abstract: | We evaluated the automated system Blood Pressure Measuring System (BPMS) developed by NASA on 277 adult males who elected to have a treadmill test as part of their annual physical. The BPMS uses acoustic transduction with a computer-assisted ECG gating to detect nonsynchronous noise. The BPMS readings were compared to pressures simultaneously measured by trained technicians. For all stages of work, BPMS readings were higher for systolic and lower for diastolic than technician readings. At peak stages of work, BPMS systolic pressures were about 20 mmHg higher than technician readings. Within each 3-min workstage, BPMS readings were found to be more inconsistent than technician readings. The standard errors of measurement for BPMS were from two to three times higher than technician values. These data showed automated blood pressure readings were significantly different than technician values and subject to more random fluctuations. These findings demonstrate the need to view exercise blood pressure measured by automated systems with caution. |