Robotic system for magnetic resonance imaging-guided focused ultrasound treatment of thyroid nodules
Authors:Antria Filippou  Nikolas Evripidou  Christakis Damianou
Institution:Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus
Abstract:

Background

Herein, a robotic system offering Magnetic Resonance-guided Focused Ultrasound (MRgFUS) therapy of thyroid nodules was developed.

Methods

The robotic system offers linear motion in 2 PC-controlled axes that navigate a 3 MHz single-element focused transducer. The system, through a C-arm structure attaches to the table of Magnetic Resonance Imaging (MRI) scanners and couples to the neck of patients lying in the supine position. The MRI compatibility of the developed system was assessed inside a 3 T scanner. Benchtop and MRI feasibility studies evaluating the heating performance of the system were executed on excised pork tissue and on homogeneous and thyroid model agar-based phantoms.

Results

The MRI compatibility of the system was successfully established. Grid sonications executed using robotic motion inflicted discrete and overlapping lesions on the excised tissue, while magnetic resonance (MR) thermometry successfully monitored thermal heating in agar-based phantoms.

Conclusions

The developed system was found to be efficient with ex-vivo evaluation. The system can perform clinical MRgFUS therapy of thyroid nodules and other shallow targets after further in-vivo evaluation.
Keywords:MRgFUS  MRI  nodules  robotic system  thyroid