The effect of contact lens-induced corneal edema on Goldmann applanation tonometry and dynamic contour tonometry |
| |
Authors: | Jong-Hyun Oh Chungkwon Yoo Yong Yeon Kim Hyo-Myung Kim Jong-Suk Song |
| |
Institution: | (1) Department of Ophthalmology, Guro Hospital, Korea University College of Medicine, 80, Guro-dong, Guro-gu, Seoul, 152–703, Republic of Korea |
| |
Abstract: | Purpose To evaluate the effect of contact lens-induced corneal edema on intraocular pressure (IOP) measurements using Goldmann applanation
tonometry (GAT) and dynamic contour tonometry (DCT) in Asian subjects.
Participants The study included 40 eyes of 20 normal volunteers with no evidence of ocular disease.
Methods Forty eyes of 20 healthy volunteers were required to wear soft contact lenses for 2 hours to induce corneal swelling. Central
corneal thickness (CCT) and IOP were measured before and immediately after contact lens wear using specular microscope, GAT,
and DCT. The IOP measurements by GAT and DCT were compared. The changes in the CCT and the IOP measurements after wearing
contact lenses were assessed.
Results The mean CCT of the 40 eyes evaluated was 532.6 ± 31.6 μm. The mean IOP was 11.78 ± 2.04 mmHg for the GAT and 14.46 ± 1.89 mmHg
for the DCT, and the difference was statistically significant (P < 0.001). After wearing contact lenses, the mean CCT was 553.2 ± 34.3 μm, which was 20.6 ± 12.9 μm greater than before wearing
them (P < 0.001). The mean IOP measurements of the GAT and DCT were decreased after wearing the contact lenses. The mean decrease
of the GAT values was 0.43 ± 1.95 mmHg, which was not statistically significant (P = 0.175). However, the mean decrease of the DCT readings, which was 0.75 ± 1.74 mm Hg, was statistically significant (P = 0.010).
Conclusion The IOP measurements with DCT were significantly higher than those with GAT in healthy Asian eyes. Although the mean IOP measurements
of both the GAT and the DCT were decreased in the edematous cornea, IOP measurements of the DCT were more affected by corneal
edema than were the GAT.
The authors have no proprietary, commercial, or financial interests in any of the products described in this study. |
| |
Keywords: | Corneal edema Dynamic contour tonometry Goldmann applanation tonometry Intraocular pressure |