How does prosthesis head size affect vibration transmission in ossiculoplasty?
Authors:Manohar Bance  Alfredo Campos  Lillian Wong  David P Morris  Rene van Wijhe
Affiliation:E.A.R. Laboratory, Dalhousie University, Halifax, Nova Scotia, Canada. m.bance@dal.ca
Abstract:OBJECTIVES: The transmission of vibrations from the tympanic membrane to the stapes footplate by an ossicular reconstruction prosthesis is affected by the size of the prosthesis head. We sought to determine if augmenting or reducing the head size of prosthesis had a systematic effect on transmission of vibrations to the stapes. STUDY DESIGN: We conducted a fresh cadaveric temporal bone middle ear study. METHODS: The incus was replaced with a prosthesis using a tympanic membrane to stapes head (TASH)-type hydroxyapatite prosthesis in nine fresh cadaveric temporal bones. Three prosthesis head sizes were created: unaltered, reduced, and augmented. Stapes vibrations were measured with a laser Doppler vibrometer in response to acoustic frequency chirps at 90 dB SPL. RESULTS: All three head size prostheses resulted in smaller stapes vibrations than the intact ear. There was no difference in the vibration transmission between the three different head sizes. All prostheses showed a vibration loss of 10 to 15 dB compared to the intact ear. CONCLUSIONS AND SIGNIFICANCE: Within the range of sizes tested, prosthesis head size had little impact on vibration transmission to the stapes footplate.
Keywords: