A Mechanical Comparison of the Compressive Force Generated by Various Headless Compression Screws and the Impact of Fracture Gap Size |
| |
Authors: | Asif M. Ilyas Jonathan M. Mahoney Brandon S. Bucklen |
| |
Affiliation: | 1.Thomas Jefferson University, Philadelphia, PA, USA;2.Globus Medical Inc, Audubon, PA, USA |
| |
Abstract: | Background: There is evidence that interfragmentary fracture gap size may affect the compression achievable with a modern headless compression screw (HCS). This mechanical study compared the compression achieved by 3 commercial HCS systems through various fracture gaps: CAPTIVATE Headless (Globus Medical, Inc, Audubon, Pennsylvania), Synthes (DePuy Synthes, Westchester, Pennsylvania), and Acumed Acutrak 2 (Acumed LLC, Hillsboro, Oregon). Methods: Screws were inserted into a custom test fixture composed of polyurethane synthetic bone foam fragments, separated by a layer of easily compressible polyurethane foam simulating a fracture gap. Compression was measured after final insertion and countersinking. The effect of the interfragmentary fracture gap size on the compression generated was also investigated. Results: The CAPTIVATE Headless 3.0 mm screw (70.1 ± 5.7 N) and the Synthes 3.0 mm screw (64.9 ± 7.3 N) achieved similar compressive forces after final countersink. Similar comparisons were found for the CAPTIVATE Headless 2.5 mm and Synthes 2.4 mm screws, and the CAPTIVATE Headless 4.0 mm and Acutrak 2 Standard screws. The final compression of the CAPTIVATE Headless 2.5 mm and Synthes 2.4 mm screws was not significantly affected when the fracture gap was doubled from 2 to 4 mm, but was reduced significantly by 95.9% with the Acutrak 2 Micro screw. Conclusion: When comparing like-sized screws, the CAPTIVATE, Synthes, and Acutrak 2 HCS systems demonstrated similar potential compressive forces. However, compared with the CAPTIVATE Headless and Synthes HCS systems, which are inserted with a compression sleeve that is not gap distance–dependent, the Acutrak 2 HCS system demonstrated less compression when the simulated fracture gap size was increased to 4 mm. |
| |
Keywords: | headless compression screw scaphoid fracture internal fixation fracture gap interfragmentary compression bone fragments compression sleeve |
|
|