Effect of cuff positioning on the accuracy of blood pressure measurement with automated electronic blood pressure monitors
Authors:Ya Li  Fang Li  Yi Li  Xiaoran Cui  Jing Li  Hua Zhi  Weidong Wang  Yanyan Sun  Wei Cui
Institution:1. Department of Cardiology, The Second Hospital of Hebei Medical University and Institute of Cardiocerebrovascular Disease of Hebei Province, Shijiazhuang China ; 2. Affiliated Hospital of Hebei University of Engineering, Handan China ; 3. The Central Hospital of Handan, Handan China
Abstract:It is recommended that the cuff should be wrapped around the upper arm with the midline of the bladder placed over the brachial artery during blood pressure (BP) measurement. However, in practice, the cuff of sphygmomanometers is often incorrectly placed. The authors aimed to assess the effect on the accuracy of BP measurement as to the placement of the cuff bladder by using oscillometric devices. Participants aged 18 years or older were enrolled. The center of the cuff bladder was placed directly over the brachial artery as the standard position (correct position), which was rotated by 90°medially (medial position), 90°laterally (lateral position), and rotated by 180°(contralateral position), respectively. The main outcomes were non‐invasive brachial BP in the four cuff positions, brachial artery pulse wave velocity, ankle‐brachial index, and invasive radial BP. Of 799 participants, 56.4% were men (60.37 ± 12.73 years), and of the 104 intensive care unit participants, 60.57% were men (57.78 ± 15.89 years). There were no significant differences in non‐invasive brachial BP among the four cuff positions (P > .1), and the mean BP differences between incorrect and standard cuff positions were within 1.0 mm Hg. BP of the incorrect positions was positively correlated with standard position (P < .001, r > .88) and showed good consistency. There was no effect on the accuracy of BP measurement as to the location of the midline of the cuff bladder by using oscillometric devices with a conventional cuff.
Keywords:blood pressure measurement  invasive  non‐invasive  oscillometric devices  position
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号