Inter- and intraobserver reliability assessment of the Cobb angle: manual versus digital measurement tools
Authors:Michaela Gstoettner  Katrin Sekyra  Nadja Walochnik  Peter Winter  Roland Wachter  Christian M. Bach
Affiliation:Department of Orthopedic Surgery, Innsbruck Medical University, Anichstrasse 35, 6020, Innsbruck, Austria.
Abstract:The objective of this study is to determine the intraobserver and interobserver reliability of end vertebra definition and Cobb angle measurement using printed and digital radiographs of 48 patients with scoliosis. The Cobb angle and the end vertebra were assessed by six observers in 48 patients with scoliosis using printed and digital radiographs. Definition of end vertebra and measurement of the Cobb angle was repeated three times with a 3 week interval. Intraclass correlation coefficients (ICC) were used to determine the interobserver and intraobserver reliabilities. 95% prediction limits for the errors in measurements are provided. For the Cobb angle a mean ICC of 0.97 was determined for intra- and interobserver reliability measurement of the printed radiographs. For the electronic radiographs a mean ICC value of 0.93 was determined for interobserver reliability and a mean ICC value of 0.96 for intraobserver reliability. Intraobserver ICC for definition of end vertebrae was 0.8 for both methods. Interobserver ICC was 0.83 for the manual and 0.74 in the digital method. One pitfall in angle measurement implies the Cobb method itself which measures in two dimensions. Until we develop a proper tri-dimensional measuring system an error is introduced. For the Cobb angle measurement the definition of end vertebrae introduces the main source of error. Digital radiography does not improve the measurement accuracy.
Keywords:Intraobserver reliability  Interobserver reliability  Scoliosis  Cobb’  s angle  Radiography
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号