首页 | 本学科首页   官方微博 | 高级检索  
检索        


Sensitivity of an automatic external defibrillator for ventricular tachyarrhythmias in patients undergoing electrophysiologic studies
Authors:M D Carlson  C S Freeman  H Garan  J N Ruskin
Institution:Clinical Electrophysiology Laboratory, Massachusetts General Hospital, Boston 02114.
Abstract:An automatic advisory external defibrillator (AED) was activated during all arrhythmias occurring at the time of 77 electrophysiologic studies in 45 patients. Sustained ventricular tachycardia (VT) occurred during 55 studies in 31 patients and nonsustained VT was induced during 10 studies in 9 patients. Ventricular fibrillation was induced 5 times in 5 patients and atrial arrhythmias with a rapid ventricular response occurred during 7 studies in 4 patients. The AED detection algorithm advised "shock" during 36 of 55 (65%) episodes of sustained VT and all 5 episodes of ventricular fibrillation. The device correctly advised "shock" for all 6 episodes of nonsustained VT that spontaneously terminated after analysis was complete. Thus, the sensitivity of the device for all ventricular arrhythmias analyzed was 47 of 66 (71%). The device recommended "shock" for all 31 episodes of rapid VT and 1 of 2 episodes of atrial fibrillation associated with systemic hypotension (32 of 33, 97%). The 19 episodes of VT for which "no shock" was advised were comparatively slow and were hemodynamically well tolerated. The device recommended "no shock" during all 21 episodes of normal sinus rhythm and all 20 episodes of rapid atrial pacing (cycle length 400 ms) analyzed. "No shock" was advised during 4 episodes of sustained narrow QRS complex supraventricular tachycardia and 4 brief episodes of nonsustained VT that terminated before analysis was complete. Thus, in this study the specificity of the AED for wide complex tachycardias was 100%.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号