首页 | 本学科首页   官方微博 | 高级检索  
     


64-slice computed tomography assessment of coronary artery stents: a phantom study
Authors:Mahnken A H  Mühlenbruch G  Seyfarth T  Flohr T  Stanzel S  Wildberger J E  Günther R W  Kuettner A
Affiliation: a Department of Diagnostic Radiology, Aachen University of Technology, Germany; Siemens Medical Solutions, Forchheim, Germany; Institute of Medical Statistics, Aachen University of Technology, Germany; and Eberhardt-Karls-University, Tuebingen, Germany
Abstract:Purpose: To compare the use of a new 64-slice computed tomography (CT) scanner with 16-slice CT in the visualization of coronary artery stent lumen.

Material and Methods: Eight different coronary artery stents, each with a diameter of 3 mm, were placed in a static chest phantom. The phantom was positioned in the CT gantry at an angle of 0° and 45° towards the z-axis and examined with both a 64-slice and a 16-slice CT scanner. Effective slice thickness was 0.6 mm with 64-slice CT and 1 mm with 16-slice CT. A reconstruction increment of 0.3 mm was applied in both scanners. Image quality was assessed visually using a 5-point grading scale. Stent diameters were measured and compared using paired Wilcoxon tests.

Results: Artificial lumen reduction was significantly less with 64-slice than with 16-slice CT. Average visible stent lumen was 53.4% using 64-slice CT and 47.5% with 16-slice MSCT. Most severe artifacts were seen in stents with radiopaque markers. Using 64-slice CT, image noise increased by approximately 30% due to thinner slice thickness.

Conclusion: Improved spatial resolution of 64-slice CT resulted in superior assessment of coronary artery stent lumen compared to 16-slice CT. However, a relevant part of the stent lumen is still not assessable with multi-slice CT.
Keywords:Coronary vessels  stents and prostheses-computed tomography CT  multidetector row-coronary vessels  CT-computed tomography (CT)  angiography
本文献已被 InformaWorld PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号