Pentacam和UBM在ICL手术患者术前中央前房深度测量的比较
引用本文:程蕾,朱冉,魏梅,王丹梅.Pentacam和UBM在ICL手术患者术前中央前房深度测量的比较[J].眼科新进展,2015,0(8):762-764.
作者姓名:程蕾  朱冉  魏梅  王丹梅
作者单位:221002 江苏省徐州市,徐州市第一人民医院眼科
摘    要:目的 探讨Pentacam与UBM在高度近视拟行可植入式角膜接触镜(implantablecollamerlens,ICL)手术患者中中央前房深度(centralanteriorchamberdepth,CACD)测量中的差异性及一致性。方法 选取高度近视拟行ICL手术患者54例(108眼),分别采用Pentacam和UBM测量该类患者CACD,采用SPSS19.0统计学软件进行分析,计量数据进行正态检验,对两种仪器所测CACD进行配对样本t检验及直线相关分析,患者屈光度(等效球镜)及CACD进行相关性分析,应用Bland-Altman法分析评价两种仪器测量方法的一致性。结果 对Pentacam所测CACD值为(3.203±0.232)mm,UBM所测CACD值为(3.156±0.233)mm,两者测量方法所测值差异有统计学意义(t=5.681,P<0.05),但根据两种测量方法相关性分析及Bland-Altman法表明两种测量方法有很好的相关性和一致性(r=0.931,P<0.05)。结论 Pentacam与UBM分别作为光学测量仪器和超声测量仪器所得测量结果相似,都可以用于高度近视拟行ICL手术患者CACD的测量,但两种仪器各有优缺点,使用时可以互相辅助应用。

关 键 词:眼前节分析仪  超声生物显微镜  前房深度可植入式隐形眼镜  人工晶状体植入术  高度近视

 Comparison of central anterior chamber depth measured by Pentacam and UBM before ICL surgery
CHENG Lei,ZHU Ran,WEI Mei,WANG Dan-Mei. Comparison of central anterior chamber depth measured by Pentacam and UBM before ICL surgery[J].Recent Advances in Ophthalmology,2015,0(8):762-764.
Authors:CHENG Lei  ZHU Ran  WEI Mei  WANG Dan-Mei
Institution:Department of Ophthalmology, the First People’s Hospital of Xuzhou . Xuzhou 221002 , Jiangsu Province . China
Abstract:Objective To discuss the difference and the consistency between Pentacam and UBM for measunng the central anterior chamber depth ( CACD) in patients of high myopia wanting to do implantable collamer lens ( ICL) surgery. Methods CACD in 54 patients (108 eyes) wanting to do ICL surgery were measured by Pentacam and UBM independently. The statistical significance of the difference between the measurements was evaluated by a paired t test and consistency was assessed by Pearson ’ s correlation and Bland-Altman analysis. Results CACD was (3. 203 + 0. 232) mm and ( 3. 156 +0. 233) mm with Pentacam and UBM , respectively , the difference between the two measurements was statistically sigruficant ( t = 5. 681 ,P < 0. 05 ) . and a high correlation and consistency were found between the two measurements by Pearson ’ s correlation and Bland-Altman analysis ( r = 0. 931 .P < 0. 05 ) . Conclusion Pentacam and UBM can be used interchangeably for CACD measurement to calculate patients of high myopia wanting to do ICL surgery.
Keywords:Pentacam  ultrasound biomicroscope  anterior chamber depth  implantable collamer lens  intraocular lens implantation  high myopia