Dosimetric consideration of individual 125I source strength measurement and a large-scale comparison of that measured with a nominal value in permanent prostate implant brachytherapy
Authors:Takahashi Yutaka  Ito Akira  Sumida Iori  Kozuka Takuyo  Gomi Kotaro  Nose Takayuki  Ito Teruo  Yamashita Takashi
Affiliation:(1) Department of Physics, The Cancer Institute, Japanese Foundation for Cancer Research, 3-10-6 Ariake, Koto-ku, Tokyo 135-8550, Japan;(2) Department of Radiation Oncology, The Cancer Institute, Ariake Hospital of Japanese Foundation for Cancer Research, Tokyo, Japan
Abstract:Purpose We investigated the difference between measured and manufacturer's nominal source strength in a large sample of a single model of 125I seeds. Physical characteristics of single seed measurement by the well-type ionization chamber were also investigated to provide dosimetric data. Materials and methods A well-type ionization chamber with a single seed holder was used to measure source strength of all 1935 125I seeds implanted in the initial 28 patients in our hospital. Physical characteristics including linearity of readings for different integral time intervals, reproducibility, isotropy, and axial positional sensitivity were assessed. To calculate the source strength, the integral charge during 30 s was measured and converted to air kerma strength. The nominal activity stated by the manufacturer was compared with the measured value. Results Linearity, reproducibility, and isotropy of the well-type ionization chamber were within 0.2%. Measured source strength was on average 2.1% (range −7.6% to +7.2%), lower than the nominal value. Standard deviation of all measured seeds was 2.0%. The maximum difference between the measured and the manufacturer's nominal source strength in each patient was −3.7%. The standard deviation averaged 1.6%. Conclusion The nominal source strength of the 125I seeds agreed well with the measured value. Our study can be helpful as guidance for individual 125I seed source strength measurement.
Keywords:125I seed  Permanent prostate implant brachytherapy  Source strength  Measurement  Well-type ionization chamber
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号