首页 | 本学科首页   官方微博 | 高级检索  
检索        


An armoured endotracheal tube obstruction in a child
Authors:Catherine Populaire  Serge Robard  Rémy Souron
Institution:1. Département d’Anesthésiologie, H?tel-Dieu, 44035, Nantes Cédex 01, France
Abstract:The case of a child whose tracheal tube became obstructed intraoperatively 30 minutes after intubation is reported. It appears that this obstruction was related to the development of bubbles within the walls which expanded upon exposure to nitrous oxide and diffusion of that gas into the bubbles. The authors want to point out the risk of gas diffusion into the walls of the tube and the possibility that repeated sterilization may enhance the development of bubbles. Les auteurs rapportent l’observation d’un enfant de sept mois dont la sonde tracheale s’est obstruée 30 minutes après l’intubation. Cette obstruction était due à l’expansion de bulles incluses dans la paroi du tube après exposition au N2O et diffusion de celui-ci à l’interieur des bulles. Ils accusent également la stérilisation répétée de la sonde armée (contre l’avis du constructeur) d’avoir favorisé le développement de ces inclusions bulleuses.
Keywords:anaesthesia: paediatric" target="_blank">anaesthesia: paediatric            equipment: tubes endotracheal" target="_blank">equipment: tubes endotracheal  complications  sterilisation
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏