Unusual cause of desynchronization in a cardiac resynchronization device |
| |
Authors: | S.S. Barold M. Steiner A. Kucher R.X. Stroobandt |
| |
Affiliation: | Florida Heart Rhythm Institute, Tampa General Circle, 33606, Tampa, FL, USA, ssbarold@aol.com. |
| |
Abstract: | This report describes a new cause of desynchronization encountered in a cardiac resynchronization device functioning in the VVIR mode. Left ventricular stimulation was inhibited when the sensor-driven rate exceeded the programmed left ventricular (LV) maximum trigger rate. With these devices, it is important to program the LV maximum trigger interval (essentially equivalent to a LV upper interval) to a value equal or faster than the sensor-driven upper rate. |
| |