首页 | 本学科首页   官方微博 | 高级检索  
     


External Quality Assessment of Blood Glucose Monitors: Problems and Practical Solutions in the External Quality Control of Point of Care Devices with Respect to the Measurement of Blood Glucose
Authors:William Graham Wood
Affiliation:Institute for the Promotion of Quality Assessment in the Medical Laboratory (INSTAND e.V.), Duesseldorf, Germany
Abstract:Point of care testing (POCT) is evolving at an ever increasing rate. This article deals mainly with the aspect of POCT for blood glucose and the problems of external quality assessment (EQA) of point of care devices (POCD). At the present time it is only possible to control precision with EQA, independent of the matrix of the test materials (synthetic polymer-base, plasma/serum, or processed whole blood). The German Federal Medical Council guidelines for laboratory performance allow an interlaboratory imprecision of ±16%. The majority of POCD fulfill these requirements. The long-term stability of results—tested by repeated distribution of the same materials over a 12-month period—is excellent, and the performance of POCD under routine condtions is usually excellent in terms of result-comparability. The problems of accuracy in terms of control materials have still to be mastered.
Keywords:accuracy   blood glucose   diabetes mellitus   evaluation   external quality assessment   external quality assessment schemes   home testing   point of care devices   point of care testing   precision   quality control