Evaluation of the impedance tonometers TGDc-01 and iCare according to the international ocular tonometer standards ISO 8612
Authors:Peter Christian Ruokonen  Thomas Schwenteck  Jörg Draeger
Affiliation:1.Department of Ophthalmology,Charité University Hospital Berlin Campus RVK,Berlin,Germany;2.Physikalisch-Technische Bundesanstalt (PTB) Berlin,Berlin,Germany;3.Department of Ophthalmology,University Medical Center Hamburg-Eppendorf,Hamburg,Germany
Abstract:
BACKGROUND: Two new tonometers have been introduced that are based on the impedance principle. Both the TGDc-01 (transpalpebral measurement) and the iCare (corneal measurement) do not require corneal anaesthesia. The present work presents an evaluation of both devices. METHODS: Comparative measurements using one of the new tonometers and applanation tonometry were performed by one investigator according to the international standard for ocular tonometer (ISO 8612). Measurements were performed on 445 eyes without corneal pathology from 243 patients. Six measurements were performed for iCare and 3 for TGDc, immediately followed by 3 applanation tonometry measurements. RESULTS: The correlation coefficient with respect to applanation tonometry was 0.81 for TGDc and 0.95 for iCare. TGDc-01 measurements showed an average deviation of 3.1 +/- 2.6 mmHg to those of Perkins applanation tonometry. The maximum difference was 28.7 mmHg below and 9.8 mmHg above the results of applanation tonometry. iCare showed an average deviation of 2.5 +/- 1.1 mmHg to Goldmann tonometry. The maximum difference was 14.5 mmHg below and 9.8 mmHg above. CONCLUSIONS: The results of both new tonometers showed a good correlation with the reference applanation tonometric methods, but the strict requirements of ISO 8612 are not fulfilled by either tonometer at present. Additionally, transpalpebral measurements with the TGDc-01 showed unacceptably high variability.
Keywords:Glaucoma  iCare  Impedance tonometer  ISO 8612  Rebound tonometer  TGDc-01  Transpalpebral tonometer
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号