iDXA,Prodigy, and DPXL Dual-Energy X-ray Absorptiometry Whole-Body Scans: A Cross-Calibration Study
Authors:Holly Hull  Qing He  John Thornton  Fahad Javed  Lynn Allen  Jack Wang  Richard N Pierson  Dympna Gallagher
Institution:1. Department of Radiological Science and Research Institute of Health Science, Yonsei University, Wonju, Gangwon 220-710, Korea;2. Korea Electrotechnology Research Institute (KERI), Ansan, Gyeonggi 426-170, Korea;1. Kuopio Musculoskeletal Research Unit (KMRU), Surgery, Institute of Clinical Medicine, University of Eastern Finland, Kuopio, Finland;2. Department of Applied Physics, University of Eastern Finland, Kuopio, Finland;3. Department of Clinical Physiology and Nuclear Medicine, Kuopio University Hospital, Kuopio, Finland;4. Department of Orthopaedics, Traumatology and Hand Surgery, Kuopio University Hospital, Kuopio, Finland;5. Institute of Clinical Medicine, Psychiatry, University of Eastern Finland, Kuopio, Finland;6. Departments of Psychiatry, Kuopio University Hospital, Kuopio, Finland;7. Department of Psychiatry, South-Savonia Hospital District, Mikkeli, Finland;8. Department of Psychiatry, North Karelia Central Hospital, Joensuu, Finland;9. Department of Psychiatry, SOSTERI, Savonlinna, Finland;10. Department of Psychiatry, SOTE, Iisalmi, Finland;11. Department of Psychiatry, Lapland Hospital District, Rovaniemi, Finland;12. Clinic of Child Psychiatry, University Hospital of Oulu, Oulu, Finland;13. Diagnostic Imaging Centre, Kuopio University Hospital, Kuopio, Finland
Abstract:PurposeTotal body fat, lean, and bone mineral content (BMC) in addition to regional fat and lean mass values for arms, legs, and trunk were compared across a pencil-beam (Lunar DPXL) and 2 fan-beam (GE Lunar Prodigy and GE Lunar iDXA) dual-energy X-ray absorptiometry (DXA) systems.MethodsSubjects were a multiethnic sample of 99 healthy adult males (47%) and females (mean ± SD: age, 46.3 ± 16.9 yr; weight, 73.4 ± 16.6 kg; height, 167.6 ± 9.7 cm; body mass index, 26.0 ± 5.2 kg/m2) who had whole-body scans performed within a 3-h period on the 3 systems. Repeated measures ANOVA was used to test the null hypothesis that the mean values for the 3 systems were equal. Translation equations between the methods were derived using regression techniques.ResultsBone mineral content (BMC): For both genders, total BMC by iDXA was lower (p  0.004) than the other systems. Lean: for males, iDXA was lower (p  0.03) than the other systems for total, trunk, and arms. For females, DPXL estimated higher (p < 0.001) lean mass compared with the other systems for total, trunk, and arms, but iDXA estimated greater legs lean mass. For both genders, all DPXL mean values were greater than Prodigy mean values (p < 0.001).Fat: in females, all the 3 systems were different from each other for total, trunk, and legs (p  0.04). For arms, DPXL and iDXA were higher than Prodigy (p < 0.0004). For males, DPXL was less (p < 0.001) for total body, trunk, and legs compared with the other 2 systems and greater than Prodigy only for arms (p < 0.0007). These data were used to derive translation equations between systems. For several measurements, the differences between systems were related to gender.ConclusionFor estimation of BMC and body composition, there was high agreement between all DXA systems (R2 = 0.85–0.99). Even so, cross-calibration equations should be used to examine data across systems to avoid erroneous conclusions.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号