超声造影与增强CT对兔肾VX2肿瘤时间-强度曲线的对比研究
引用本文:曾功君,高云华,谭开彬,刘卫金,刘平,卞爱娜. 超声造影与增强CT对兔肾VX2肿瘤时间-强度曲线的对比研究[J]. 中国超声医学杂志, 2005, 21(9): 653-655
作者姓名:曾功君  高云华  谭开彬  刘卫金  刘平  卞爱娜
作者单位:1. 400037,重庆市,解放军第三军医大学新桥医院超声科
2. 400037,重庆市,解放军第三军医大学新桥医院放射科
基金项目:重庆市科委攻关项目(渝科发计字No.[2001]66号文),国家自然科学基金资助项目(.30200265)
摘    要:目的通过对比声学造影与增强CT的时间-强度曲线,初步探讨本科室自制的脂质体超声造影剂的显像机理。方法采用低机械指数,常规超声频率,对14只左肾种植有VX2肿瘤的家兔进行观察,造影前显示肿瘤的最大切面,注射超声造影剂后固定探头切面10min,得到声学造影的灰阶时间-强度曲线,在48h内做CT增强的时间-密度曲线。结果肾肿瘤和肾实质声学造影和增强CT的时间曲线达峰值时间无统计学差异(P>0.05),曲线下降斜率有显著性差异(P<0.05)。结论CT和超声造影剂在肾脏内药物代谢动力学不同,通过肾脏后的CT造影剂经肾小球毛细血管滤过由肾脏排泄,而本实验室自制的超声造影剂可在体内多次循环灌注肾脏,是血池显像剂。

关 键 词:肾VX2肿瘤  声学造影  增强CT  时间-强度曲线
收稿时间:2005-03-31
修稿时间:2005-03-31

Comparison Study on Time-intensity Curves of Renal VX2 Carcinoma in Ultrasound Contrast and Contrast-enhanced CT
Zeng Gongjun, Gao Yunhua, Tan Kaibin, et al. Comparison Study on Time-intensity Curves of Renal VX2 Carcinoma in Ultrasound Contrast and Contrast-enhanced CT[J]. Chinese Journal of Ultrasound in Medicine, 2005, 21(9): 653-655
Authors:Zeng Gongjun   Gao Yunhua   Tan Kaibin   et al
Affiliation:Depart. Ultrasound, Xinqiao Hospital, Chongqing 400037 China
Abstract:Objective To investigate the imaging princ iple of the self-made liposome contrast agent (Zhifuxian),the comparison of the time-intensities acquired by the ultrasound contrast imaging and the CT contra st imaging was done. Methods A total of 14 rabbits planted VX2 carcinoma in the left kidney were involved in the study before the injection of the contrast agent th e maximum imaging acquired with the low mechanic index (MI) diagnosis ultrasound ,then the imaging was stabled after the injection of the liposome contrast agent ,the first 10min images were recorded and the time-intensity curves were acquir ed according to the images and the time-intensity of the helical CT contrast im aging were also acquired in 48hs.Results As to the time to pea k there was no statistically significant between the ultrasound contrast imaging and the helical CT contrast imaging,but for the slope of the descending curve t here was statistically significant. Conclusions The pharmacokinetics of the ultrasound contrast age nts and the CT contrast agent was different,CT contrast agents is excreted thro ugh the renal corpuscles,whereas the self-made ultrasound agents could repeated ly perfuse the renal corpuscles.
Keywords:VX2 tumor in kidney  Contrast-enhanced ultrasound   Contrast enhancement CT  Time-intensity curve.