首页 | 本学科首页   官方微博 | 高级检索  
     


Changes in knee motion over the first 3 years with a mobile-bearing prosthesis
Authors:Russo Alessandro  Montagna Luisa  Bragonzoni Laura  Visani Andrea  Marcacci Maurilio
Affiliation:Biomechanics Laboratory at Istituti Ortopedici Rizzoli, Via di Barbiano 1/10 40136, Bologna, Italy. a.russo@biomec.ior.it
Abstract:The present work aims to evaluate the mobility of the polyethylene bearing in 25 patients implanted with a mobile-bearing PCL-retaining TKR (Interax ISA, Striker Howmedica) under in vivo, weight-bearing conditions over 3 years' follow-up. We compared the bearing motion in response to specific tests in two consecutive follow-ups with Roentgen Stereophotogrammetric Analysis (RSA). Results showed that polyethylene continued to displace on the tibial baseplate over time: longitudinal rotations and medio-lateral translations tended to increase at the second follow-up (mean increases 0.7+/-2.5 degrees and 0.34+/-1.08 mm, respectively), while sagittal translations did not show any relevant change. Longitudinal rotations and AP translations were preserved when present at both follow-ups. Nevertheless, no statistical evidence of repeatable pattern of motion was found, since bearings frequently reversed their pattern of motion between the two follow-ups. The unconstrained polyethylene mobility resulted in an unrepeatable, erratic pattern of bearing motion intra-patient and over time. The changes in knee motion were found not to affect the clinical results over the first 3 years of follow-up. As in a fixed-bearing design, mobile-bearing design was not able to reproduce normal knee kinematics.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号