首页 | 本学科首页   官方微博 | 高级检索  
     


A new tool for benchmarking cardiovascular fluoroscopes.
Authors:S Balter  F A Heupler  P J Lin  M H Wondrow
Affiliation:Lenox Hill Hospital, New York, New York 10021, USA. sbalter@aol.com
Abstract:This article reports the status of a new cardiovascular fluoroscopy benchmarking phantom. A joint working group of the Society for Cardiac Angiography and Interventions (SCA&I) and the National Electrical Manufacturers Association (NEMA) developed the phantom. The device was adopted as NEMA standard XR 21-2000, "Characteristics of and Test Procedures for a Phantom to Benchmark Cardiac Fluoroscopic and Photographic Performance," in August 2000. The test ensemble includes imaging field geometry, spatial resolution, low-contrast iodine detectability, working thickness range, visibility of moving targets, and phantom entrance dose. The phantom tests systems under conditions simulating normal clinical use for fluoroscopically guided invasive and interventional procedures. Test procedures rely on trained human observers.