经导管Amplatzer堵塞器治疗动脉导管未闭的评价
引用本文:周爱卿,高伟.经导管Amplatzer堵塞器治疗动脉导管未闭的评价[J].临床儿科杂志,1999,17(4):195-197.
作者姓名:周爱卿  高伟
作者单位:上海第二医科大学附属新华医院
摘    要:为研究评价新的自膨性Amplatzer堵塞器关闭动脉导管未闭(PDA)的疗效,于1998年8月至1999年2月应用Amplatzer堵塞器关闭突兀 ,年龄0.8~11岁(平均3.9岁),体重7~39kg(平均15.8kg),PDA最狭处直径2.3~6.43mm(平均3.5mm)。应用6F长鞘经股静脉插至降主动脉递送堵塞器,术后10~15分钟作主动脉造影评价即职效。关闭术后24小时、1月、3月、6月

关 键 词:动脉导管未闭  Amplatzer  堵塞器  介入疗法

Assessment for transcatheter closure of patent ductus arteriosus using a new Amplatzer duct occluder
Zhou Aiqing,Gao Wei,Yu Zhiqing,Wang Rongfa,Li Feng,Zhu Ming.Assessment for transcatheter closure of patent ductus arteriosus using a new Amplatzer duct occluder[J].The Journal of Clinical Pediatrics,1999,17(4):195-197.
Authors:Zhou Aiqing  Gao Wei  Yu Zhiqing  Wang Rongfa  Li Feng  Zhu Ming
Institution:Zhou Aiqing,Gao Wei,Yu Zhiqing,Wang Rongfa,Li Feng,Zhu Ming. Xinhua Hospital affiliated to Shanghai Second Medical University,Shanghai Institute for Pediatric Research,Shanghai 200092
Abstract:The aim of the present study was to assess the effect for the catheter closure of patent ductus artenosus (PDA) using a new self-expandable device, Amplatzer duct occluder (ADO). Twenty-six female and 6 male children accepted the transcatheter closure of PDA using the new device at a mean age of 3. 9 years old (ranged 0. 6-11) and weight of 15. 8kg (ranged 7--39kg). The average of PDA diameter at its narrowest segment was 3. 5 mm (ranged 2- 3-6. 43mm). A 6F long sheath was used for delivery of the ADO. At 10-15 minutes after the release, a descending aortogram was performed to assess the immediate results. Follow-up evaluation for the main pulmonary artery with color flow mapping was at 24 hours, 1, 3 and 6 months after closure, respectively. The results showed that all 32 cases succeeded for the device placement. The angiography confirmed that patent ductus arteriosus in 30 patients had been closed completely and only 2 cases remained a small residual shunt. One of two cases had complete closure within 24 hours and another did at one month confirmed by color Doppler. The mean time of the procedure was 42. 5 minutes and of the fluoroscopy was 6. Iminutes. No any complications were found. It is suggested that the new ADO is simple, safe and few complications for the catheter closure of the moderate or large sized PDA. However, further clinical trials and follow-up evaluation will be required.
Keywords:patent ductus arteriosus transcatheter closure Amplatzer duct occluder