首页 | 本学科首页   官方微博 | 高级检索  
检索        

不同检测系统总胆固醇测定结果的偏倚分析
引用本文:刘军,鄢斌,王植雄.不同检测系统总胆固醇测定结果的偏倚分析[J].实用医学杂志,2006,22(1):88-90.
作者姓名:刘军  鄢斌  王植雄
作者单位:434020,湖北省荆州市中心医院医学检验部
摘    要:目的:通过对不同检测系统进行方法对比分析,探讨各系统之间检测总胆固醇(代HOL)结果是否具有可比性.方法:参照NCCLS的要求,以可溯源的检测系统为目标检测系统,均采用酶法测定TCHOL对本院3个不同检测系统进行朗道质控物(水平2和水平3)各测定21次和测定新鲜血清标本40份.结果:朗道质控物和新鲜血清标本代HOL测定经随机区组设计资料的方差分析,各检测系统间的差异均具有显著性(P<0.001).各检测系统测定新鲜血清标本TCHOL可靠性系数α分别为0.9977、0.9990,各系统间的相关系数均大于0.975.各检测系统测定代HOL的精密度CV均小于5%,以可溯源的检测系统1为目标检测系统,临床可接受性能评价TCHOL均超出临床接受范围.结论:3个不同检测系统测定TCHOL结果不具有可比性,当同一实验室同一检验项目存在两个或两个以上检测系统时,应进行结果的对比和偏倚评估.

关 键 词:胆固醇  检测系统  偏倚  
收稿时间:2005-08-30
修稿时间:2005-08-30

Bias analysis on TCHOL results among different kinds of detection system
LIU Jun,YAN Bin,WANG Zhi-xiong.Bias analysis on TCHOL results among different kinds of detection system[J].The Journal of Practical Medicine,2006,22(1):88-90.
Authors:LIU Jun  YAN Bin  WANG Zhi-xiong
Institution:Department of Clinical Laboratory, Jinzhou Center Hospital, Jingzhou 434020, China
Abstract:Objective Discuss the comparability of total cholesterol(TCHOL) results among different detection system. Methods 3 different kinds of biochemistry detection systems were used to detect TCHOL concentration in 2 levels of Randox quality controls and 40 clinical sera according to EP9- A file. The collected data were dealt with statistics analysis. Results Analysis of variance showed TCHOL results from different control and patients sera have significant difference in different detection system (P< 0.001). The correlation coefficient of each detection system were all above 0.975 and the reliability coefficient of them was 0.9977 and 0.9990. Target system 1 was regarded as standard detection system and evaluated the acceptability of other detection system.. As a result, they were not accepted for TCHOL by clinic assessment. Conclusion The comparability of TCHOL results in 3 kinds of detection system did not accord to evaluation of clinical acceptability. If there are more than two systems for the same test, it is necessary to do method comparison and bias estimation in order to insure the comparability.
Keywords:detection system Total Cholesterol Bias
本文献已被 CNKI 维普 万方数据 等数据库收录!
点击此处可从《实用医学杂志》浏览原始摘要信息
点击此处可从《实用医学杂志》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号