首页 | 本学科首页   官方微博 | 高级检索  
检索        


A new hearing screening system for preschool children
Authors:Wenjin Wu  Jingrong Lü  Yun Li  Anna Chi Shan Kam  Michael Chi Fai Tong  Zhiwu Huang  Hao Wu
Institution:1. Department of Otolaryngology-Head and Neck Surgery, Xinhua Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200092, China;2. Ear Institute, Shanghai Jiaotong University School of Medicine, Shanghai 200092, China;3. Department of Otorhinolaryngology, Head and Neck Surgery and Institute of Human Communicative Research, The Chinese University of Hong Kong, Hong Kong, China
Abstract:

Objectives

This study aimed to investigate the practical application of Smart Hearing, a new hearing screening system for preschool children.

Methods

The screening system was applied to 6288 preschool children. The system auto-tested hearing thresholds at three frequencies: 1 kHz, 2 kHz, and 4 kHz; a 30 dB hearing level (HL) was the critical intensity for passing. Children with positive results were referred for audiological evaluation (pure tone audiometry, tympanometry and distortion product otoacoustic emissions assessment, etc.). To evaluate the test accuracy, 312 children (5%) were randomly selected to receive audiology assessment.

Results

In this study, 582 children (9.3%) tested positive in the screening, and the referral rate of the four age groups from 3 to 6 years old was 18.8%, 11.9%, 6.5% and 4.0%, respectively. A total of 463 children underwent audiological assessment, of which 12 cases (1.91‰; 95% CI: 0.83‰, 2.99‰) were diagnosed with permanent hearing loss, and 75 cases (1.19%; 95% CI: 0.92%, 1.46%) were diagnosed with temporary conductive hearing loss. No mixed hearing loss was found in this study. The specificity of the system was 92.6% and the sensitivity was only 37.5%.

Conclusions

This screening system is suitable for the universal hearing screening of preschool children above 4 years old, and further improvements of the system are needed to increase its sensitivity.
Keywords:CI  confidence interval  dBA  decibels A-weighted  HL  hearing level  OAE  otoacoustic emissions  PTA  pure tone audiometry
本文献已被 ScienceDirect 等数据库收录!