In vitro evaluation of 56 coronary artery stents by 256-slice multi-detector coronary CT |
| |
Authors: | Henning Steen,Florian André ,Grigorios Korosoglou,Dirk Mueller,Waldemar Hosch,Hans-Ulrich Kauczor,Evangelos Giannitsis,Hugo A. Katus |
| |
Affiliation: | aUniversity of Heidelberg, Department of Cardiology, Im Neuenheimer Feld 410, Heidelberg 69120, Germany;bPhilips GmbH Healthcare Division, Luebeckertordamm 5, Hamburg 20099, Germany;cUniversity of Heidelberg, Department of Diagnostic and Interventional Radiology, Im Neuenheimer Feld 410, Heidelberg 69120, Germany |
| |
Abstract: | ObjectiveWe sought to investigate stent lumen visibility of 56 coronary stents with the newest 256-multi-slice-CT (256-MDCT) technology for different reconstruction algorithms in an in vitro model.BackgroundEarly identification of in-stent restenosis (ISR) is important to avoid recurrent ischemia and prevent acute myocardial infarction (AMI). Since angiography has the disadvantage of high costs and its invasiveness, MDCT could be a convenient and safe non-invasive alternative for detection of ISR.Material and methodsPercentages of in-stent lumen diameter and in-stent signal attenuation (measured as contrast-to-noise ratio (CNR)) of 56 coronary stents (group A ≤2.5 mm; group B = 2.75–3.0 mm; group C = 3.5–4.0 mm) were evaluated in a coronary vessel in vitro phantom (iodine-filled plastic tubes) employing four different reconstruction algorithms (XCD, CC, CD, XCB) on a novel 256-MDCT (Philips-iCT, collimation = 128 mm × 0.625 mm; rotation time = 270 ms; tube current = 800 mA s with 120 kV). Analysis was conducted with the semi-automatical full-width-at-half-maximum (FWHM) method. P-values <0.05 were regarded statistically significant.ResultsIn-stent lumen diameter >60% for group C stents was significantly larger and CNR was significantly lower (both p < 0.05) for sharp kernels (CD; XCD) when compared to groups A/B. The FWHM-method showed significantly smaller in-stent lumen diameter (p < 0.05) when compared to the manual method.Conclusion256-MDCT could potentially be employed for clinical assessment of stent patency in stents >3.0 mm when analysed with cardio-dedicated sharp kernels, although clinical studies corroborating this claim should be performed. However, stents ≤3.0 mm reconstructed by soft kernels revealed insufficient in-stent lumen visualisation and should not be used in clinical practice.Further improvements in spatial and temporal image resolution as well as reductions of radiation exposure and image noise have to be accomplished for the ambitious goal of characterising both CT coronary artery anatomy and in-stent lumen. |
| |
Keywords: | Abbreviations: AMI, acute myocardial infarction BMS, bare metal stent CAD, coronary artery disease DES, drug eluting stent ISR, in-stent restenosis MDCT, multi-detector CT PCI, percutaneous coronary intervention |
本文献已被 ScienceDirect 等数据库收录! |
|