双探头SPECT符合线路衰减校正对图像分辨率及计数影响的模型研究
引用本文:耿建华,陈英茂,陈盛祖,郑容,张雯杰,田嘉禾. 双探头SPECT符合线路衰减校正对图像分辨率及计数影响的模型研究[J]. 中国医学影像技术, 2005, 21(1): 124-126
作者姓名:耿建华  陈英茂  陈盛祖  郑容  张雯杰  田嘉禾
作者单位:1. 中国医学科学院肿瘤医院核医学科,北京,100021
2. 中国人民解放军总医院核医学科,北京,100853
摘    要:目的研究衰减校正对双探头符合成像系统分辨率及图像计数的影响.方法采用GE Hawkeye 双探头符合成像设备,对分辨率模型成像,用X线透射扫描进行衰减校正.计算有、无衰减校正时线源扩展函数的半高宽(FWHM)及线源峰值计数.结果有衰减校正时FWHM显著高于无衰减校正时FWHM,其差值在中心处最大,且线源计数大于无衰减校正时计数,越靠中心其差值越大.结论衰减校正在补偿计数衰减的同时,会使分辨率降低.

关 键 词:衰减校正  分辨率  双探头符合成像
文章编号:1003-3289(2005)01-0124-03
收稿时间:2004-08-12
修稿时间:2004-08-12

Influence of attenuation correction on imaging resolution and counts for dual-head SPECT coincidence imaging: a phantom study
GENG Jian-hu,CHEN Ying-mao,CHEN Sheng-zu,ZHENG Rong,ZHANG Wen-jie and TIAN Jia-he. Influence of attenuation correction on imaging resolution and counts for dual-head SPECT coincidence imaging: a phantom study[J]. Chinese Journal of Medical Imaging Technology, 2005, 21(1): 124-126
Authors:GENG Jian-hu  CHEN Ying-mao  CHEN Sheng-zu  ZHENG Rong  ZHANG Wen-jie  TIAN Jia-he
Affiliation:Department of Nuclear Medicine, Cancer Hospital, PUMC&CAMS, Beijing 100021, China;Department of Nuclear Medicine, General Hospital of PLA, Beijing 100853, China;Department of Nuclear Medicine, Cancer Hospital, PUMC&CAMS, Beijing 100021, China;Department of Nuclear Medicine, Cancer Hospital, PUMC&CAMS, Beijing 100021, China;Department of Nuclear Medicine, Cancer Hospital, PUMC&CAMS, Beijing 100021, China;Department of Nuclear Medicine, General Hospital of PLA, Beijing 100853, China
Abstract:Objective To investigate the influence of attenuation correction (AC) on dual head coincidence imaging resolution and counts. Methods The resolution phantom was imaged using GE Hawkeye dual head coincidence detection equipment. AC was performed with X ray transmission scan. The full width at half maximum (FWHM) of the line resource spread function and the peek pixel counts of the line resource were calculated with and without AC. Results The FWHM with AC was significant larger than that without AC, and the smaller the distance from the center, the larger the difference. The peek pixel counts with AC were more than those without AC, and the counts difference was maximal at center and minimal at the edge of the field of the view. Conclusion AC can compensate the count attenuation but also impair the imaging resolution.
Keywords:Attenuation correction  Resolution  Dual-head coincidence imaging