首页 | 本学科首页   官方微博 | 高级检索  
检索        

验光仪顶焦度测量方法的不确定度分析
引用本文:王敬涛,贾晓航.验光仪顶焦度测量方法的不确定度分析[J].医疗装备,2014(1):11-13.
作者姓名:王敬涛  贾晓航
作者单位:浙江省医疗器械检验院,杭州310009
摘    要:按照行业标准的要求,提出了验光仪顶焦度测量误差的具体测量方法,即:用验光仪测量标准模拟眼,每个模拟眼测7次,计算平均值,平均值与模拟眼的标定值比较,其差值即为验光仪的测量误差。并分析了该方法的不确定度。

关 键 词:验光仪  顶焦度测量方法  不确定度

Analysis of uncertainty of measurement method for vertex power of eye refractometers
Institution:WANG Jing - tao, JIA Xiao - hang (Zhejiang Institute for the Control of Medical Device, Hangzhou, 310009)
Abstract:This paper is to set up a method to measure the vertex power' s error of the eye refraetometer according to the trade standard. That is to measure the standard model eyes by eye refractometer, each one is measured seven times, and we calculate the ar- ithmetical average value. Arithmetical average value minus normal value of the model eye, the difference is the error. It also analyze the uncertainty of the method.
Keywords:Eye refractometers  Measuring method for vertex power  Uncertainty
本文献已被 CNKI 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏