首页 | 本学科首页   官方微博 | 高级检索  
检索        


Low-friction hydrophilic surface for medical devices
Authors:S Nagaoka  R Akashi
Institution:Basic Research Laboratories, Toray Industries, Inc., Kanagawa, Japan.
Abstract:A hydrophilic polymer surface was developed exhibiting excellent low frictional property, namely slipperiness, when in contact with water or physiological fluid due to the reaction of epoxy-containing poly(vinyl pyrrolidone) with the polyamino compound formed on the surface of the substrate. Epoxy-containing poly(vinyl pyrrolidone) was obtained by the copolymerization of vinyl pyrrolidone as a hydrophilic component, glycidyl acrylate as a binding component to the substrate, and vinyl acetate to preserve the strength of the coating layer. The surface friction coefficient depends on the molecular weight of the coated hydrophilic copolymer. It was demonstrated that a molecular weight of 400,000 or more is essential to achieve excellent low surface friction. Using rabbit models, polyurethane catheters, both with and without the hydrophilic low friction coating, were evaluated for surface friction coefficient and blood compatibility. As a result, in the case of coated catheters, no lesions of the intima of the blood vessels and no thrombus formations on the surfaces of the catheters were observed. However, the non-coated catheters injured the intima of the blood vessels and severe thrombus formation was found on their surfaces.