首页 | 本学科首页   官方微博 | 高级检索  
     


Measurement of ventilation and respiratory mechanics during continuous positive airway pressure (CPAP) treatment in infants
Authors:B Andréasson  M Lindroth  N W Svenningsen  B Drefeldt  G I J?nsson  L Niklason  B Jonson
Affiliation:Department of Paediatrics, University Hospital, Lund, Sweden.
Abstract:A new method has been evaluated for measuring ventilation and lung mechanics in spontaneously breathing infants by means of a face chamber. Airway flow is measured with a pneumotachograph inserted between the face chamber and a stable pressure source. Oesophageal pressure is measured via a water-filled oesophageal catheter. The method is suitable for use in conjunction with continuous positive airway pressure (CPAP) treatment in neonatal intensive care. A flat frequency response curve up to 15 Hz for the two measuring systems (i.e., airway flow and oesophageal pressure), and a time shift between the two respective signals of less than 2 msec are prerequisites for correct evaluation of respiratory mechanics. In preterm infants with chest distortion, the inhomogeneity of pleural pressure affects the significance of resistance and compliance values, as calculated from oesophageal pressure. Supra-diaphragmatic pressure variations reflect the resistive and elastic load on the diaphragm exerted by the lungs and thorax. Thus, oesophageal pressure is still useful in studies of respiratory mechanics in preterm infants.