首页 | 本学科首页   官方微博 | 高级检索  
检索        


Tensile strength testing for resorbable mesh fixation systems in laparoscopic ventral hernia repair
Authors:Emmelie Reynvoet  Frederik Berrevoet  Filip De Somer  Griet Vercauteren  Ingrid Vanoverbeke  Koen Chiers  Roberto Troisi
Institution:Department of General and Hepatobiliary Surgery and Liver Transplantation Service, Ghent University Hospital and Medical School, De Pintelaan 185, 2K12-IC, 9000, Ghent, Belgium, emmelie.reynvoet@ugent.be.
Abstract:

Background

In an attempt to improve patient outcome and quality of life after laparoscopic ventral hernia repair, resorbable fixation devices have been developed to allow adequate mesh fixation while minimizing accompanying side-effects as tack erosion and adhesion formation.

Materials and methods

In experimental set-up, 24 pigs were treated by laparoscopic mesh placement. Two different meshes (PP/ORC and PP/ePTFE) and four fixation devices were evaluated: a 6.4?mm poly(D,L)-lactide pushpin (tack I), a 6.8?mm poly(D,L)-lactide with blunt tip (tack II), a 4.1?mm poly(glycolide-co-L-lactide) (tack III) and one titanium tack (control tack). A first group of animals (n?=?12) was euthanized after 2?weeks survival and a second group (n?=?12) after 6?months. At euthanasia, a relaparoscopy was performed to assess adhesion formation followed by laparotomy with excision of the entire abdominal wall. Tensile strength of the individual fixation systems was tested with the use of a tensiometer by measuring the force to pull the tack out of the mesh. Additionally, the foreign body reaction to the fixation systems was evaluated histologically as was their potential degradation.

Results

At 2?weeks the tensile strength was significantly higher for the control tack (31.98?N/cm2) compared to the resorbable devices. Except for tack II, the tensile strength was higher when the devices were fixed in a PP/ePTFE mesh compared to the PP/ORC mesh. After 6?months only tack III was completely resorbed, while tack I (9.292?N/cm2) had the lowest tensile strength. At this time-point similar tensile strength was observed for both tack II (29.56?N/cm2) and the control tack (27.77?N/cm2). Adhesions seem to be more depending on the type of mesh, in favor of PP/ePTFE.

Conclusion

At long term, the 4.1?mm poly(glycolide-co-L-lactide) tack was the only tack completely resorbed while the 6.8?mm poly(D,L)-lactide tack with blunt tip reached equal strengths to the permanent tack.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!