Accuracy of a 3D temporal scanning system for gait analysis: Comparative with a marker-based photogrammetry system
Institution:1. Instituto de Biomecánica de Valencia, Universitat Politècnica de València, edifici 9C. Camí de Vera, s/n, 46022 València, Spain;2. Instituto Universitario de Automática e Informática Industrial, Universitat Politècnica de València, edifici 1F. Camí de Vera, s/n, 46022 València, Spain
Abstract:BackgroundCombining the accuracy of marker-based stereophotogrammetry and the usability and comfort of markerless human movement analysis is a difficult challenge. 3D temporal scanners are a promising solution, since they provide moving meshes with thousands of vertices that can be used to analyze human movements.Research questionCan a 3D temporal scanner be used as a markerless system for gait analysis with the same accuracy as traditional, marker-based stereophotogrammetry systems?MethodsA comparative study was carried out using a 3D temporal scanner synchronized with a marker-based stereophotogrammetry system. Two gait cycles of twelve healthy adults were measured simultaneously, extracting the positions of key anatomical points from both systems, and using them to analyze the 3D kinematics of the pelvis, right hip and knee joints. Measurement differences of marker positions and joint angles were described by their root mean square. A t-test was performed to rule out instrumental errors, and an F-test to evaluate the amplifications of marker position errors in dynamic conditions.ResultsThe differences in 3D landmark positions were between 1.9 and 2.4 mm in the reference pose. Marker position errors were significantly increased during motion in the medial-lateral and vertical directions. The angle relative errors were between 3% and 43% of the range of motion, with the greatest difference being observed in hip axial rotation.SignificanceThe differences in the results obtained between the 3D temporal scanner and the marker-based system were smaller than the usual errors due to lack of accuracy in the manual positioning of markers on anatomical landmarks and to soft-tissue artefacts. That level of accuracy is greater than other markerless systems, and proves that such technology is a good alternative to traditional, marker-based motion capture.
Keywords:3D temporal scanner  Stereophotogrammetry  Markerless  Gait analysis  Landmark position error
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号