Migration of the Duraloc cup after 5 years |
| |
Authors: | Christoph Stihsen Christof Pabinger Roman Radl Peter Rehak Reinhard Windhager |
| |
Institution: | (1) Department of Orthopaedic Surgery, Medical University of Graz, Auenbruggerplatz 5, 8036 Graz, Austria;(2) Med-Center Waltendorf, Plüddemanngasse 39, 8010 Graz, Austria;(3) Department of Surgery, Medical University of Graz, Auenbruggerplatz 29, 8036 Graz, Austria |
| |
Abstract: | The Duraloc cup is a frequently used metal-backed, porous-coated, hemispherical, press-fit acetabular component. Published data on loosening rates are contradictory. In this study we investigated migration patterns with computer-assisted Einzel-Bild-Roentgen-Analyse (EBRA) of 67 Duraloc 100 cups. Cup migration and clinical scores were analysed over a 5-year follow-up period. Median total migration of the Duraloc 100 cup was 1.21 mm at 5 years. Seventy-five percent of implants were radiologically stable at 2 years and 90% at 4 years. One cup loosened aseptically at 60 months, requiring revision. Cup diameters > or = 54 mm migrated significantly more than cups < 54 mm in diameter (p = 0.029 at 4 years). There was a significant correlation between high polyethylene wear and further migrating cups within the first post-operative year (p = 0.035 at 12 months). Our analysis revealed significantly higher wear in males (p = 0.029 at 4 years). Radiological loosening at two years could be calculated using receiver-operating characteristic curve analysis, and 1.2 mm as an adequate threshold value (sensitivity = 100%, specificity = 89%). |
| |