人脑血氧水平依赖信号有源仿真和成像
引用本文:李洪宇,贾春涛,陈迢,邱建峰. 人脑血氧水平依赖信号有源仿真和成像[J]. 中国医学影像技术, 2020, 36(4): 590-595
作者姓名:李洪宇  贾春涛  陈迢  邱建峰
作者单位:山东科技大学海洋科学与工程学院, 山东 青岛 266590,山东科技大学海洋科学与工程学院, 山东 青岛 266590,山东第一医科大学(山东省医学科学院)医学工程技术研究中心, 山东 泰安 271016,山东第一医科大学(山东省医学科学院)医学工程技术研究中心, 山东 泰安 271016
基金项目:山东省泰山学者项目(ts201712065)。
摘    要:目的 对人脑血氧水平依赖(BOLD)信号进行仿真并成像,以之作为金标准和标准信号,基于信号成像灰度幅值变化评价脑功能磁共振成像(fMRI)质量和检出率。方法 设计一种能有源模拟人脑BOLD的激发电路,实现不同电流幅值、不同频率、不同占空比的稳定矩形波输出,在时间域内仿真人脑BOLD信号。结果 3台不同型号、不同场强MR皆有稳定的矩形波输入,且检出率一致。结论 同一设备对同一刺激模式检出率相近,可作为已知且确定信号的参考标准,定量分析MR系统对BOLD信号的检出率和一致性。

关 键 词:磁共振成像  血氧水平依赖  矩形波输出
收稿时间:2019-05-15
修稿时间:2020-01-02

Active simulation and imaging of human brain blood oxygenation level dependent signal
LI Hongyu,JIA Chuntao,CHEN Tiao and QIU Jianfeng. Active simulation and imaging of human brain blood oxygenation level dependent signal[J]. Chinese Journal of Medical Imaging Technology, 2020, 36(4): 590-595
Authors:LI Hongyu  JIA Chuntao  CHEN Tiao  QIU Jianfeng
Affiliation:Ocean Science and Engineering College, Shandong University of Science and Technology, Qingdao 266590, China,Ocean Science and Engineering College, Shandong University of Science and Technology, Qingdao 266590, China,Medical Engineering and Technology Research Center, Shandong First Medical University, Tai''an 271016, China and Medical Engineering and Technology Research Center, Shandong First Medical University, Tai''an 271016, China
Abstract:Objective To active simulation and imaging of human brain blood oxygenation level dependent (BOLD) signal, taken which as gold standard and standard signal to detect the quality and detection rate of brain MRI functional magnetic resonance imaging (fMRI) based on the changes of gray amplitude of BOLD signals during imaging. Methods A new excitation circuit that could simulate the BOLD of the human brain was used to achieve the stable rectangular wave output of different current amplitude, different frequency and different space ratio, and the BOLD signal of the human brain in the time domain was simulated. Results Stable rectangular wave input and consistent detection rate were obtained with 3 different MR devices with different field strength. Conclusion The same device has similar detection rate for the design of the same stimulus model, which can be used as the reference standard of the known and determined signal to analyze the detection rate and consistency of BOLD signal.
Keywords:magnetic resonance imaging  blood oxygenation level dependent  rectangular wave output