Pulsed Doppler: determination of diameter, blood flow velocity, and volumic flow of brachial artery in man |
| |
Authors: | J A Levenson P A Peronneau A Simon M E Safar |
| |
Abstract: | A pulsed Doppler velocimeter suitable for the determination of blood flow velocity and volumic flow in peripheral arteries is described. The apparatus has two main characteristics: an adjustable range-gated time system and a double transducer probe. The error in the determination of the angle between the ultrasound beam and flow of blood with this apparatus was less than 2%, and overestimation of the arterial diameter due to the sample volume size did not exceed 0.035 +/- 0.015 cm. The apparatus was used to determine diameter, blood flow velocity and volumic flow of the brachial artery of 22 healthy men. The values were respectively 0.440 +/- 0.010 cm, 9.15 +/- 1.01 cm.s-1 and 85 +/- 10 cm3.min-1. Administration of intravenous nitroglycerin significantly increased the arterial diameter (p less than 0.001) without any significant change in volumic flow. The described pulsed Doppler velocimeter provides an accurate noninvasive method for determining volumic flow in peripheral arteries in clinical investigation and cardiovascular pharmacology. |
| |