3D radial sampling and 3D affine transform‐based respiratory motion correction technique for free‐breathing whole‐heart coronary MRA with 100% imaging efficiency
Authors:Himanshu Bhat  Lan Ge  Sonia Nielles‐Vallespin  Sven Zuehlsdorff  Debiao Li
Affiliation:1. Departments of Radiology and Biomedical Engineering, Northwestern University, Chicago, IL, USA;2. Siemens AG Medical, MREA‐CV, Erlangen, Germany;3. Siemens Medical Solutions USA, Inc., Chicago, IL, USA;4. Cedars‐Sinai Medical Center, University of California, Los Angeles, CA, USA
Abstract:The navigator gating and slice tracking approach currently used for respiratory motion compensation during free‐breathing coronary magnetic resonance angiography (MRA) has low imaging efficiency (typically 30–50%), resulting in long imaging times. In this work, a novel respiratory motion correction technique with 100% scan efficiency was developed for free‐breathing whole‐heart coronary MRA. The navigator signal was used as a reference respiratory signal to segment the data into six bins. 3D projection reconstruction k‐space sampling was used for data acquisition and enabled reconstruction of low resolution images within each respiratory bin. The motion between bins was estimated by image registration with a 3D affine transform. The data from the different respiratory bins was retrospectively combined after motion correction to produce the final image. The proposed method was compared with a traditional navigator gating approach in nine healthy subjects. The proposed technique acquired whole‐heart coronary MRA with 1.0 mm3 isotropic spatial resolution in a scan time of 6.8 ± 0.9 min, compared with 16.2 ± 2.8 min for the navigator gating approach. The image quality scores, and length, diameter and sharpness of the right coronary artery (RCA), left anterior descending coronary artery (LAD), and left circumflex coronary artery (LCX) were similar for both approaches (P > 0.05 for all), but the proposed technique reduced scan time by a factor of 2.5. Magn Reson Med, 2011. © 2011 Wiley‐Liss, Inc.
Keywords:coronary MRA  3D PR k‐space sampling  3D affine transform  respiratory motion correction
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号