首页 | 本学科首页   官方微博 | 高级检索  
     


Scattered radiation level during videofluoroscopy for swallowing study
Authors:Chan C B  Chan L K  Lam H S
Affiliation:Department of Radiology, Kwong Wah Hospital, Kowloon, Hong Kong. changcb@ha.org.hk
Abstract:AIM: To evaluate the scattered radiation exposure to the surroundings during videofluoroscopy for swallowing study (VFSS). MATERIALS AND METHODS: Scattered radiation exposure was measured using an ion chamber survey meter for 17 adult patients undergoing videofluoroscopy for swallowing study. The cumulative dose area product of each case was also recorded. Data were presented as mean +/- standard deviation. RESULTS: The scattered radiation exposure at a distance of 150 cm from the patient and the dose area product recorded were 149 +/- 78 microR (range 42-308 microR) and 842 +/- 544 cGy.cm(2) (range 258-2151 cGy.cm(2)), respectively, for a single study of 18 +/- 6 minutes. A formula was then derived for estimating the scattered radiation dose to muscle tissue at an arbitrary distance based on the accumulated dose area product. With this formula, the mean scattered radiation dose to naked muscle tissue of the surrounding people at a distance of 30-100 cm from the patient were estimated to be 33.68-3.03 microSv respectively. CONCLUSION: The scattered radiation detriment associated with videofluoroscopy for swallowing study was well within acceptable levels.
Keywords:scattered radiation   dose area product   videofluoroscopy for swallowing study
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号