Long-Term Results of Nervous Tissue Alterations Caused by Epineurial Electrode Application: An Experimental Study in Rat Sciatic Nerve
Authors:RUPERT KOLLER,WERNER GIRSCH,CHRISTIAN LIEGL,,HELMUT CRUDER,,JÜ  RGEN HOLLE,UDO LOSERT,WINFRIED MAYR,HERWIG THOMA
Affiliation:Second Surgical Clinic, University of Vienna, Austria.
Abstract:In order to evaluate the long-term effects of epineurial electrode application for functional electrical stimulation (FES) the left sciatic nerve of seven rats was exposed. Four ring-shaped stainless steel wire electrodes were sutured to the epineurium of each nerve in the same manner as performed clinically for carrousel stimulation in man. The nerves were reexposed 1 year after implantation and the stimulation threshold to obtain a tetanic contraction in the lower limb was determined for each electrode. Afterwards the animals were sacrificed. The electrodes were excised and cross sections of the sciatic nerve directly at site of the electrodes, 2-mm proximal and 2-mm distal to them were harvested for histologic and planimetric assessment of nerve lesions. The area of damaged neural tissue was expressed as a percentage of the total cross-sectional area within the perineural sheath. The sciatic nerves of the right side served as controls. The values for the stimulation thresholds ranged between 0.1 and 1.0 mA (mean 0.43 mA). By morphometric examination five of seven nerves were seen altered, the altered areas captured between 1% and 4.8% of the total cross-sectional area of the nerves within the perineural sheath. Besides two specimens, all altered nerve segments exhibited distinct signs of nerve fiber regeneration. The clinical implications of the results for long-term electrical stimulation, such as phrenic pacing, are discussed.
Keywords:electrode    epineurial    functional electrical stimulation    diaphragm pacing    nerve alterations    morphometry
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号