Comparative Efficacy of Subcutaneous Mesh and Plate Electrodes for Nonthoracotomy Canine Defibrillation
Authors:AKIHIKO NOGAMI,ATSUSHI TAKAHASHI,,JUNICHI NITTA,&dagger  ,KAZUTAKA AONUMA,,YOSHITO IESAKA,,MICHIAKI HIROE,FUMIAKI MARUMO,MASAYASU HIRAOKA&dagger  
Affiliation:Second Department of Internal Medicine, Tokyo Medical and Dental University, Japan.
Abstract:To determine the optimal configuration for the subcutaneous placement of electrodes for the performance of ventricular defibrillation without thoracotomy, internal defibrillation using four different subcutaneous electrodes was performed in 13 anesthetized dogs (7-12 Kg, mean +/- SD: 9.2 +/- 1.5 Kg). An electrode (7 cm2) was positioned transvenously in the superior vena cava with the following electrodes randomly implanted subcutaneously on the left chest: small mesh electrode (14 cm2), large mesh electrode (28 cm2), small titanium plate electrode (14 cm2), and large plate electrode (28 cm2). Ventricular fibrillation was induced by applying alternating current; a monophasic defibrillation wave was administered between the superior vena cava and the subcutaneous electrodes 10 seconds later. The energy level associated with a 50% successful defibrillation, as predicted by logistic regression analysis, was defined as the ED50. After the completion of the defibrillation protocol using the four subcutaneous electrodes, the small mesh electrode was sutured to the epicardium and the ED50 measurements were repeated. Energy ED50s were lower when the superior vena cava electrode was used as the cathode rather than as the anode. Of the subcutaneous electrodes, the large plate electrode showed the lowest energy ED50 (3.3 +/- 0.9 joules). The plate electrodes had lower energy ED50s than the mesh electrodes, and the large electrode had a lower energy ED50 than the small electrodes. Using the epicardium electrode, transient arrhythmias and ST elevation were observed following successful defibrillation; however, no arrhythmias or ST-T changes were observed following defibrillation using the subcutaneous electrodes.(ABSTRACT TRUNCATED AT 250 WORDS)
Keywords:nonthoracofomy internal defibrillation    subcutaneous plate electrode    electrode size
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号