Artifacts seen during CT pelvimetry: implications for digital systems with scanning beams
Authors:Brody, AS   Saks, BJ   Field, DR   Skinner, SR   Capra, RE
Abstract:During computed tomography (CT) pelvimetry of two pregnant women, bony abnormalities of the fetuses were noted on the scout images that were not confirmed at delivery. To explore the cause of these artifacts, specimen long bones were manipulated in various ways during CT scout imaging. Artifacts like those seen during in vivo imaging were found to be caused by motion of the object. The CT scout view is an example of an image produced by a digital system that uses a scanning beam. This type of digital system is being used for several types of body imaging including screening for scoliosis and chest radiography. Attention to motion artifacts should decrease diagnostic errors and aid further development of these systems.