Variomatrix--a newly developed transesophageal echocardiography probe with a rotating matrix biplane transducer. Technological aspects and initial clinical experience
Authors:Omoto R  Kyo S  Matsumura M  Yamada E  Matsunaka T
Institution:Department of Surgery, Saitama Medical School, Japan.
Abstract:The recently developed multiplane transesophageal echocardiography (TEE) probes cannot provide real-time synchronous biplane imaging. We have developed a new "Variomatrix" probe that combines the advantages of the rotating mechanism of the multiplane probe with the capability of providing simultaneous, side-by-side imaging. Its advantages are: (1) easy manipulation to obtain the optimal planes; (2) avoidance of excessive movement of the probe inside the esophagus; (3) assessment of oblique planes that are not viewed with the usual biplane probe; (4) better understanding of the continuity of structures by the simultaneous, side-by-side biplane, rotating images, providing clearer three-dimensional conception; and (5) shorter examination time by providing a 180 degrees imaging range by rotating the transducer just 90 degrees .
Keywords:multiplane transesophageal echocardiography  VariomatrixTM transesophageal echocardiography
本文献已被 PubMed 等数据库收录!