Abstract:The improvements achievable by introducing individual tissue compensators in photon therapy of the breast were assessed. In 37 patients the dose ranged from +15% to -10% of the mid target dose using combinations of wedge filters and beam weights alone. With a tissue compensator the dose ranged from +4% to -11% provided that allowance was made for lung attenuation. A megavoltage imaging system is a potential source of the X-ray transmission data which can provide a basis for the calculation of thickness of the compensator.