SL-IC放射治疗模拟机高压电缆击穿原因分析
引用本文:杨秋权.SL-IC放射治疗模拟机高压电缆击穿原因分析[J].中国医学装备,2011,8(7):98-100.
作者姓名:杨秋权
作者单位:揭阳市人民医院放疗科,广东,揭阳,522000
摘    要:目的:解决我科放射治疗模拟机高压电缆几个月就被击穿的现象;方法:通过对SL-IC放射治疗模拟机高压电路的分析,总结引起该机高压电缆易被击穿的原因;结果:在最终选用厂家提供的合格的高压发生器后使故障解决;结论:放射治疗模拟机高压电路故障隐蔽,要和厂家密切联系,并经过反复试验和认真总结,快速完成机器维护工作,更好地服务临床。

关 键 词:X线机  高压电路  故障  维修

Analysis of reasons for breakdown in high voltage cable of SL-IC radiotherapy simulator
Yang Qiu-quan.Analysis of reasons for breakdown in high voltage cable of SL-IC radiotherapy simulator[J].China Medical Equipment,2011,8(7):98-100.
Authors:Yang Qiu-quan
Institution:Yang Qiu-quan Department of Radiotherapy,People's Hospital of Jieyang City,Jieyang 522000,China.
Abstract:Objective:To solve the problem of breakdown in high voltage electrocircuit of radiotherapy simulator in a couple of months in our department. Methods:The high voltage electrocircuit of radiotherapy simulator was analyzed to determine reasons for its breakdown. Results:After selecting an eligible high voltage generator provided by manufacturer,the failure was solved finally. Conclusion:As for solution of the failure about the high voltage electrocircuit of radiotherapy simulator,we must keep in touch with th...
Keywords:X-ray machine  High voltage electrocircuit  Maintenance  