首页 | 本学科首页   官方微博 | 高级检索  
     


Prediction of Optimal Atrioventricular Delay in Patients with Implanted DDD Pacemakers
Authors:TOSHIYUKI ISHIKAWA  SHINICHI SUMITA  KAZUO KIMURA  MIYAKO KIKUCHI  MASAMI KOSUGE  NAOMITSU KUJI  TSUTOMU ENDO  TERUYASU SUGANO  TOMOHIKO SIGEMASA  IZUMI KOBAYASHI  OSAMU TOCHIKUBO  TAKASHI USUI
Affiliation:Second Department of Internal Medicine, Yokohama City University Urafune Hospital, Yokohama, Japan.
Abstract:In patients with an implanted DDD pacemaker (PM), the atrial contribution may be interrupted by too short an atrioventricular (AV) delay, and filling time may be shortened by too long an AV delay. The AV delay at which the end of the A wave on transmitral flow coincides with complete closure of the mitral valve may be optimal. The subjects were 15 patients [70.3+/-12.3 (SD) years old] with an implanted DDD PM. Cardiac output (CO) and pulmonary capillary wedge pressure (PCWP) were measured by Swan-Ganz catheter. Transmitral flow was recorded by pulsed Doppler echocardiography. AV delay was prolonged stepwise by 25 msc. When the AV delay was set at 155+/-26 ms, the end of the A wave coincided with complete closure of the mitral valve. When the AV delay was prolonged 25, 50, 75, and 100 ms from this AV delay, the interval between the end of the A wave and complete closure of mitral the valve was prolonged 16+/-5, 39+/-6, 65+/-4 and 88+/-5 ms, respectively (r = 0.97, P<0.0001) and diastolic mitral regurgitation was observed during this period. Thus, the optimal AV delay may be predicted as follows: the slightly prolonged AV delay minus the interval between the end of the A wave and complete closure of the mitral valve. When the AV delay was set at 215 ms, there was a significant positive correlation between the predicted optimal AV delay (166+/-23 ms) and the optimal AV delay (CO: 161+/-26 msec, r = 0.93, P<0.0001, PCWP: 161+/-28 msec, r = 0.95, P<0.0001). In conclusion, optimal AV delay can be predicted by this simple formula: slightly prolonged AV delay minus the interval between end of A wave and complete closure of mitral valve at the AV delay setting.
Keywords:Doppler echocardiography    diastolic mitral regurgitation    PQ interval    DDD pacemaker    atrioventricular block
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号